| Specify whether the connected pumps are operated via the bipolar or the unipolar output of the 12-bit-DAC-card. The voltage range of unipolar outputs is from 0 to 5V or 10V. For bipolar outputs, the range is from -5/10V to +5/10V. If possible, use only the unipolar outputs. Dionex recommends using the unipolar outputs, as only this guarantees 0V at the 12-DAC-card output when turning on the pump and the computer! |