| DGP-3600 
		 and LPG-3400 (RS, SD or BM, not 
		 XRS) pumps only: Specify 
		 the settings for pump stroke synchronization when one of the pumps 
		 listed above is operated with an OAS-3x00TXRS 
		 autosampler. In addition, 
		 you can specify a shutdown action for the Corona detector, in 
		 case the pump flow decreases to 0.   OAS-3x00TXRS 
		 Sync – Select this check box 
		 to enable pump stroke synchronization for the pump. When using a dual DGP-3600RS/SD/BM 
		 pump, you can specify whether synchronization is to be performed 
		 for the left or right pump unit. Clear 
		 the check box to disable pump stroke synchronization with the 
		 autosampler. For more information about how to use the pump with 
		 an OAS-3x00TXRS, refer to  UltiMate 3000 Series OAS-3x00TXRS Autosampler: 
		 Installation.   Emergency 
		 Shutdown (Coulochem/ECD-RS) – Selecting 
		 this check box enables a Coulochem 
		 or ECD-RS detector to automatically start a Shutdown procedure 
		 when the pump flow becomes 0. This option is enabled by default. When 
		 using a dual DGP-3600RS/SD/BM pump, you can specify whether synchronization 
		 is to be performed for the left or right pump unit. Emergency Shutdown (Corona) 
		 - Selecting this check box enables a HPLC pump to automatically 
		 start a Shutdown procedure when the gas flow becomes 0. This option 
		 is enabled by default. |