Page:  
Autosampler
Enter the configuration of the installed autosampler:
 Tip:
 Tip:  
If you have selected Get configuration from the instrument on the Select configuration mode page, all options on the Autosampler page are disabled. Clicking Next> takes you to the following wizard pages; click Finish to complete the installation. When the installation is complete, double-click the device name and determine the rack manually. The description of the options below refers to manual configuration.
| Option | Description | 
| 
 | 
 | 
| Type | Select L7200 if an autosampler is installed with your HPLC system. | 
| 
 | 
 | 
| Device Name: | Accept the name under which the instrument is identified in the installation environment (default setting: Sampler; also, refer to Naming Devices...). If you do not accept the default name, you may have to edit the PGM Files and re-link the corresponding control of the control panel. (Also, refer to How to: Controlling Devices from the Control Panel  | 
| 
 | 
 | 
| Rack | Enter the rack parameters: | 
| 
 | 
 | 
| Type | Select the autosampler type: User (Please define the rack geometry.)   | 
| 
 | 
 | |||
| UserNx | Field | Description | Min | Max | 
| Nx | Number of vials per row | 1 | 20 | |
| Ny | Number of vials per column | 1 | 20 | |
| X1 | Distance from the Home position to the center of the nearest vial on the x-axis [mm] | 0 | 156 | |
| Y1 | Distance from the Home position to the center of the nearest vial on the y-axis [mm] | 0 | 107 | |
| X2 | Distance from the Home position to the farthest vial on the x-axis [mm] | 0 | 156 | |
| Y2 | Distance from the Home position to the farthest vial on the y-axis [mm] | 0 | 107 | |
| Z 
 | Needle height [mm] above the sampling location when the needle is in Home position. | 0 | 57 | |
| 
 | 
 | |||
| Operation pattern | Define the processing order for the vials in the rack: 1 (Row-by-Row): | |||
| 
 | 
 | |||
| 
 | 
 | |||
| Syringe | Select the volume (100 or 500 µl) of the installed syringe. | |||
| 
 | 
 | |||
| Sample Loop | Enter the volume of your sample loop (in µl). | |||
A wizard guides you through the initial installation of the instrument:
| <Back | Clicking <Back returns you to the previous wizard page. | 
| Next> | Clicking Next> takes you to the next wizard page. | 
Also, refer to Installing and Controlling Third-Party Devices