Installing and Controlling Third-Party 
 Devices
 Gilson 333 and 334 Pumps: Installation
 
 Gilson 333 and 334 Pumps: Installation
	
		| Device Type: | Pumps | 
	
		| Device Driver: | Gilson: 333/334 Pump | 
	
		| Supported Hardware 
		 Options: | 
			Operation as preparative master/slave 
			 system where the 333 pump is the master pump and up to two 
			 334 pumps are the dependent pumps (slaves)Stand-alone operation of the 
			 333 pumpAdditional option: Solvent 
			 Valve Block (SVB) | 
 
What is required?
	
		| License: | Timebase 
		 Class 3 Select About 
		 Chromeleon on the Help 
		 menu of the Chromeleon client to check whether this license class 
		 is available. | 
	
		| Connection: | Required 
		 component for the Gilson 333 pump: Required component for the Gilson 334 
		 pump: | 
	
		| HardwarePrerequisites:
 | 
			A multi-serial 8-fold PCI interface 
			 card (8-RS-232 Multi-COM card, Thermo Scientific part no. 
			 5906.2095)A free RS-232 port on the Multi-COM 
			 card. (Note: Connection via other RS-232 ports, e.g., the 
			 COM port of the UCI, 
			 is not possible.) | 
	
		| Control 
		 Panel: | GIL333_334.pan | 
Hardware Installation
	
		| Device Connection: | The 
		 device driver supports the preparative Gilson 333 pump as a master 
		 pump and up to two Gilson 334 pumps as dependent pumps (slaves) 
		 for the delivery of other solvents. First, connect the Gilson 333 pump 
		 to the PC:  Connect the GSIOC from Controller 
		 port on the pump to a serial interface port on the PC, using the 
		 Gilson 605 GSIOC RS-232 adapter cable. (For cable details, refer 
		 to
  Gilson 
		 605 GSIOC RS-232 Adapter Cable in the Appendix 
		 of the Hardware Installation 
		 section. The Gilson part number for the adapter cable is 360784192. 
		 The cable is not available from Thermo Fisher Scientific.) Next, connect the Gilson 334 pump to 
		 the 333 pump:  Use the Gilson GSIOC cable to connect the GSIOC 
		 from Controller port on the 334 pump to the GSIOC 
		 to Slave port on the Gilson 333 pump. (The GSIOC cable 
		 is shipped with the Gilson 334 pump; the cable is not available 
		 from Thermo Fisher Scientific.)
 | 
	
		| Device Settings: |  Tip:
 For 
		 more information about the settings described below, refer to 
		 the user's guide for the instrument. Number of Pumps Enter the number of Gilson 334 slave 
		 pumps on the pump's control panel. On the Edit 
		 menu, select Config and 
		 then select Hard. GSIOC 
		 Addresses Assign a unique GSIOC address to each 
		 pump. For the Gilson 333 pump, enter the GSIOC address on the 
		 pump's control panel. For stand-alone operation of the Gilson 
		 333 pump, enter any address between 0 and 63. To operate the pump 
		 as the master pump, the 
		 address must be 1. For the Gilson 334 slave pump(s), the 
		 address(es) must be 2 
		 for solvent B (and 3 for 
		 solvent C). Set the address via the switch on the pump's rear 
		 panel. Baud 
		 Rates The Gilson 333 pump uses a baud rate 
		 of 19200. Set the baud rate on the pump's control panel. The baud 
		 rate for the Gilson 334 pump can be set to 19200 or to EXTERNAL. 
		 Set the baud rate via the switches on the pump's rear panel. Solvent 
		 Valve Block (SVB) When you use a Solvent Valve Block 
		 (SVB), set the SVB manually in the device configuration of the 
		 master pump. To set the SVB on the control panel of the master 
		 pump, follow the description in the pump's User's Guide. When 
		 connecting to Chromeleon, the driver checks the SVB configuration 
		 of the pumps and compares it to the settings in the Server Configuration 
		 program. If the settings do not match, an error message appears 
		 and the connect action is aborted. | 
	
		| Restrictions: | Support 
		 is not available for the following features of the Gilson 333 
		 and 334 pumps: 
			Control of the Gilson 30x pumps 
			 as injection unitsUV detector controlTwin pump option, i.e., doubling 
			 the flow rate by controlling two identical pump systems on 
			 the same GSIOC busUse of input signals to trigger 
			 specific actionsExecution of service actions 
			 (calibration, priming, leak tests, etc.) while the pump is 
			 controlled by Chromeleon. Execute these actions on the pump's 
			 control panel when the pump is not controlled by Chromeleon. | 
 
Software Installation
Install the Device Driver in the 
 Chromeleon Server Configuration. (For more information about the program, 
 refer to Software Installation and Communication 
  The 
 Server Configuration 
 Program.)
 The 
 Server Configuration 
 Program.)
	- Start 
	 the Chromeleon server. 
- Start 
	 the Server Configuration 
	 program. 
- Add 
	 the pump to the timebase. Select Add Device on the Edit 
	 or context menu, and then select Gilson 
	 from the Manufacturers list 
	 box and 333/334 Pump from 
	 the Devices list box. 
- The 
	 following settings are required: 
On the General 
 tab page, enter a name for the pump. The default name is Pump. Select 
 a free COM port for the GSIOC communication. In addition, verify that 
 the Demo Mode is disabled.
On the Pump 
 Configuration (Gilson 333/334 Pumps) tab page, select the GSIOC addresses 
 and the installed head types. In addition, specify whether a solvent valve 
 block is installed.
On the Flow 
 and press. limits tab page, determine the allowed flow and pressure 
 range for the pump.
On the Relays 
 tab page, select all relays that you want to control.
The Error 
 Levels tab page classifies the severity of any errors that occur. 
 It is generally not necessary to change the default settings.
	- To 
	 control the installed Gilson pump, connect the GIL333_334.pan 
	 Control Panel or any 
	 especially developed control panel in the Chromeleon client to the 
	 corresponding timebase. 
Application
Set the following communications parameters: 
 19200 Baud, 8 Bit, 2 Stop, Even Parity, and no Handshake. Use these parameters 
 also when forming gradients or operating two or three pumps.
 Tip:
 Tip: 
  
When Chromeleon connects to the pump, the 
 connect procedure lasts about 30 to 40 seconds. After that, Chromeleon 
 can control the pump. This is logged in the Audit 
 Trail.
When Chromeleon controls the pump, the pump's 
 control panel is locked. The CM remote 
 control message on the control panel indicates that the pump is 
 connected to Chromeleon
Chromeleon monitors the pressure. That is why 
 pressure limits set on the pump are ignored when Chromeleon controls the 
 instrument. If the pressure is below the lower limit, the pump will stop 
 after some time.
For an overview of special commands and properties 
 for these pumps, refer to Commands and 
 Tips for Third-Party Devices  Gilson 331 to 334 Pumps in the User Help section.
 Gilson 331 to 334 Pumps in the User Help section.
 Note:
 Note: 
  
If you use the pumps in both local and remote 
 operation, please note that Chromeleon overwrites the internal method 
 files 18-20.
Further Information
 
 Tip:
 Tip:
For more installation details, refer to 
 the user's guide for the pump.
 
For an overview of the different Gilson instruments 
 for which device drivers are available in Chromeleon, refer to  Gilson.
 Gilson.