Installing and Controlling Third-Party 
 Devices
 Varian 3600 GC: Installation
 
 Varian 3600 GC: Installation
| Device Type: | Gas chromatograph | 
| Device Driver: | Varian: 3600 Gas Chromatograph | 
| Supported Hardware Options: | Components 
 Detector 
 | 
What is required?
| License: | Timebase Class 1 | 
| Connection: | |
| Hardware | The following hardware and firmware versions (or previous versions) are required: CPU, 
		 IBDH ASSY 03-925367-33 REV 1, SCHEM 03-925370-00 with the 
		 following EPROMs: 03-910692-01 
		 REV-A FCB4, An A/D converter is required for data acquisition. A UCI Universal Chromatography Interface, a Dionex UI20 (Dionex part no. 46017), or any other A/D converter that can be controlled by Chromeleon, e.g., PE Nelson Boxes (including the appropriate signal cables per channel) may be used. 
 Make sure that the sampler is always connected directly to the gas chromatograph. A free RS-232 port is required. | 
| Varian3400.pan | 
Hardware Installation
| Device Connection: | Use 
		 a special 9-pin serial cable (Dionex part no. 8914.0145) to connect 
		 the Varian 3600 GC. (For cable details, refer to  
 Do not attempt to connect the cable directly with the PC. Make sure that you connect that end of the cable to the PC that is correspondingly marked. In addition, one Varian signal cable 
		 (Dionex part no. 8911.3600) is required per detector. (For cable 
		 details, refer to  For data acquisition, an A/D cable is required, also. The cable depends on the A/D converter used, e.g., a UCI Universal Chromatography Interface). | 
| Device Settings: | The following communication parameters are used: 9600 baud, 7 data bits, 1 stop bit, even parity, and no protocol. 
 | 
| Restrictions: | Chromeleon does not support temperature ramps for the injector. 
 Dionex advises against manual input on the device. Chromeleon cannot read all settings from the device. Therefore, manual settings on the device may lead to errors in the program. | 
Software Installation
Install the Device 
 Driver in the Chromeleon Server Configuration. (For more information 
 about the program, refer to Software 
 Installation and Communication  The 
 Server Configuration 
 Program.)
 The 
 Server Configuration 
 Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the gas chromatograph to the timebase. Select Add Device on the Edit or context menu, and then select Varian from the Manufacturers list box and 3600 Gas Chromatograph from the Devices list box.
The following settings are required:
On the Components tab page, specify the components that are installed.
On the Oven tab page, specify the limits for the controllable temperature range of the column oven.
On the InjectorA/B tab page, enable temperature control for the injector oven and define the controllable temperature range.
On the Detector A/B tab page, enable temperature control for the detector oven and define the controllable temperature range. In addition, specify the detector type.
On the Aux tab page, enable temperature control of the auxiliary device and define the controllable temperature range.
Open the Signal Configuration dialog box from the Signals tab page by double-clicking the respective channel. Assign a channel to the signal and name the signal in the Signal Name field according to the selected detector type.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the gas chromatograph, connect the Varian3400.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
The User Help 
 section provides information about the commands and properties supported 
 by the 3600 GC; refer to Commands and 
 Tips for Third-Party Devices  Varian.
 Varian.
Troubleshooting
 Tips:
 Tips:
If a sample sequence includes a blank run, several Can’t set property error messages may be displayed at the beginning of the next sample. These messages can be ignored.
If Detector B Flameout appears on the device and/or in Chromeleon, but if the detector B is not used or not installed: Disconnect the device in Chromeleon, press the Reset key on the device, and reconnect the device to Chromeleon. In addition, delete the following command in the PGM File:
GC_2.Autozero.
For more information about troubleshooting, 
 refer to Commands and Tips for Third-Party 
 Devices  Varian 3400/3600 GCs: Troubleshooting 
 in the User Help section.
 Varian 3400/3600 GCs: Troubleshooting 
 in the User Help section.
If you have any special problems, please contact Dionex Service.
Further Information
 Tip:
 Tip:
For more installation details, refer to the manual for the instrument.