Calibration Examples
 Calibration Modes for External Calibration
 Calibration Modes for External Calibration
The Calibration Mode determines the standard samples that are used for calibrating specific unknown samples. The following calibration modes are available:
| Calibration Mode | Description | 
| Total | All standard samples of the current sequence | 
| Group | Grouping calibration | 
| Additional | Additional standard samples for the samples that appear later in the sequence. | 
| Bracketed | Unknown samples are "bracketed" by standard samples. | 
| Fixed | Specific standard samples (also from other sequences) | 
For a schematic representation of the Group, Additional, and Bracketed calibration modes, refer to the example below:

For more information about the different calibration modes including calibration curve calculation, refer to the topics below:
 Note:
 Note:
Please note that the procedure is described for the Total mode. All other examples describe the differences between the calibration modes only.