Dionex Ion Chromatography Systems
 Dionex DX-120 Ion Chromatograph
 Dionex DX-120 Ion Chromatograph
The following commands are available for control of DX-120 operating functions (please note that the display Filter level determines which commands are displayed):
| Command | Description | 
| Column | If the DX-120 is in Column Mode, the Column command sets the flow path to column A or column B. Refer to  | 
| Controlled AC | Turns the AC power outlet on the DX-120 rear panel on and off. This provides on off control of an external accessory connected to the outlet. Refer to the DX-120 operator’s manual for cabling instructions. | 
| Sets the rate at which Chromeleon collects digital data points from the detector. | |
| Eluent | If the DX-120 is in Eluent Mode, the Eluent command selects the eluent reservoir (A or B). Refer to  | 
| Eluent Pressure | Turns the pressure to the eluent reservoir(s) on and off. | 
| Pressure Unit | Select the units of pressure to use (psi or MPa). | 
| Pump | Turns the pump on and off. | 
| SRS | Turns the current supplied to the Self-Regenerating Suppressor (SRS) on and off. |