In addition to the standard commands available for all detectors (see  Dionex Detectors), Dionex electrochemical detectors provide special commands (see the table below; please note that the display Filter level determines which commands are displayed). The Data_Collection_Rate command and the analog output commands are available for all Dionex electrochemical detectors in any operating mode. The Mode command is available only for the ED40/ED50/ED50A detectors. Other special commands depend on the selected operating mode.
 Dionex Detectors), Dionex electrochemical detectors provide special commands (see the table below; please note that the display Filter level determines which commands are displayed). The Data_Collection_Rate command and the analog output commands are available for all Dionex electrochemical detectors in any operating mode. The Mode command is available only for the ED40/ED50/ED50A detectors. Other special commands depend on the selected operating mode.
| Command | Description | 
| Sets the rate at which Chromeleon collects digital data points from the detector. | |
| Mode | (Available only for the ED40/ED50/ED50A) Select the operating mode: conductivity, integrated amperometry, or DC amperometry. | 
Analog Output Commands (All Modes)
| Command | Description | 
| Sets the full-scale analog out voltage. | |
| Sends a positive pulse to the detector’s analog output as an event marker. | |
| Sets the offset for the analog output signal. | |
| Switches the analog output polarity (+/-). | |
| Sets the range of a full-scale recorder response. | |
| Sets how quickly the detector responds to a change in signal. | 
Special Commands (Conductivity Mode)
The following commands are available for conductivity detectors (CD20/CD25/CD25A, ED40/ED50/ED50A, or IC20/IC25/IC25A):
| Command | Description | 
| DS3_Temperature | Sets the DS3 temperature. | 
| SRS_Current | For CD20/CD25, ED40/ED50, IC20/IC25 only: Sets the current supplied to the SRS. | 
| Suppressor_Current | For CD25A, ED50A, and IC25A only: Sets the current supplied to the suppressor. | 
| Suppressor_Type | For CD25A, ED50A, and IC25A only: Sets the installed Suppressor. | 
| Sets the temperature compensation factor, which is used to stabilize conductivity readings. | 
Special Commands (DC Amperometry Mode)
The following commands are available for the ED40/ED50/ED50A detectors in DC amperometry mode:
| Command | Description | 
| Cell | Turns the cell on and off. | 
| DC_Voltage | Sets the fixed potential applied to the working electrode, in the range of -2.04V to +2.04V. | 
| Electrode | Select the type of electrode installed. | 
Special Commands (Integrated Amperometry Mode)
The following commands are available for the ED40/ED50/ED50A detectors in integrated amperometry mode:
| Command | Description | 
| Cell | Turns the cell on and off. | 
| Electrode | Select the reference electrode type. | 
| Define a waveform (a plot of potential vs. time). | 
Calibration Commands (For CD25A, ED50A, and IC25A Only)
| Command | Description | 
| LeakDet | Performs the leak detector calibration. | 
| CondCell | Performs the conductivity cell calibration | 
| DownloadCalibration | Downloads the selected set of calibration values to the detector. | 
| DownloadCalibration Parameter | Selects the set of calibration values (Current, Previous, or Factory) to be downloaded. | 
Diagnostic Commands (For CD25A, ED50A, and IC25A Only)
| Command | Description | 
| LeakDetTest | Performs a test of the leak detector. | 
| LeakDetTestDate | Reports the date of the leak detector test. | 
| LeakDetTestResult | Reports the result (pass or fail) of the leak detector test. |