Dionex Detectors
 Dionex PDA/PDA-100/PDA-3000 Photodiode Array Detectors
 Dionex PDA/PDA-100/PDA-3000 Photodiode Array Detectors
In addition to the standard commands available for all detectors (see  Dionex Detectors), the PDA, PDA-100, and PDA-3000 detectors provide several additional control commands (please note that the display Filter level determines which commands are displayed).
 Dionex Detectors), the PDA, PDA-100, and PDA-3000 detectors provide several additional control commands (please note that the display Filter level determines which commands are displayed).
UV_VIS Channel Commands
| Command | Description | 
| Sets the bandwidth (the wavelength range at which the chromatogram is measured). | |
| Sets the offset for the analog output signal. | |
| Sets the range of a full-scale recorder response (analog output). | |
| Sets the bandwidth of the reference wavelength (if one is selected). If RefWavelength is Off, the RefBandwidth setting has no effect. | |
| Sets the reference wavelength. | 
3DFIELD Channel Commands
| Command | Description | 
| Determines how many wavelengths are averaged when collecting 3D data. Bunching (averaging) the signals of adjacent wavelengths will reduce the amount of 3D data stored. | |
| MaxWavelength | Sets the maximum wavelength of the 3D field. | 
| MinWavelength | Sets the minimum wavelength of the 3D field. | 
| RefBandwidth | Sets the bandwidth of the reference wavelength (if one is selected). If RefWavelength is Off, the RefBandwidth setting has no effect. | 
| RefWavelength | Sets the reference bandwidth. | 
Other Special Detector Commands
| Command | Description | 
| AmbientTemp | Reports the temperature outside the optical assembly. | 
| AZ_Done | Reports the status of an autozero command (Busy or Done). | 
| Sets the rate at which Chromeleon collects digital data points from the detector. | |
| Sets the amount of increased light transmission from autozero (negative absorption) to which the detector responds. This permits peaks to be seen even when the eluent decreases absorbance. | |
| OpticalBenchTemp | Reports the temperature of the optical bench. | 
| Sets how quickly the detector responds to a change in signal. | |
| Sets the TTL input signal mode. | |
| Turns the deuterium lamp on and off. | |
| UVLampPowerOnSetting | Selects whether the deuterium lamp is turned on when the detector power is turned on. The default setting is On. | 
| Turns the tungsten lamp on and off. | |
| VisLampPowerOnSetting | Selects whether the tungsten lamp is turned on when the detector power is turned on. The default setting is On. | 
UV Calibration Commands
| Command | Description | 
| DownloadCalibration | Downloads the selected set of calibration values to the detector. | 
| DownloadCalibration Parameter | Selects the set of calibration values (Current, Previous, or Factory) to be downloaded. | 
| FilterWheel | Calibrates the filter wheel (PDA-100 (DX-LAN) only). | 
| Intensity | Calibrates the intensity of the deuterium lamp. | 
| LeakDet | Performs the leak detector calibration (PDA-100 (DX-LAN) only). | 
| WavelengthCal | Performs a wavelength calibration. | 
| WavelenthCalDate | Reports the date of the wavelength calibration. | 
| WavelenthCalResult | Reports the result (pass or fail) of the wavelength calibration. | 
UV Diagnostic Commands
| Command | Description | 
| LeakDetTest | Performs a test of the leak detector (PDA-100 (DX-LAN) only). | 
| LeakDetTestDate | Reports the date of the leak detector test. | 
| LeakDetTestResult | Reports the result (pass or fail) of the leak detector test. | 
| WavelengthVerTest | Performs wavelength verification via the Holmium Oxide Filter. | 
| WavelengthVerTestDate | Reports the date of the wavelength verification. | 
| WavelengthVerTestResult | Reports the result (pass or fail) of the wavelength verification. | 
System Wellness and Predictive Performance commands and properties
 Tip:
 Tip:  
The Chromeleon version and the detector type determine which commands and properties are available for checking the detector wellness and performance.
| Command/Property | Description | 
| Reports the deuterium lamp's intensity at 254 nm in counts/second. | |
| ModuleAge | Reports (or resets) the number of hours the detector has been on. | 
| Qualification | Provides information about detector qualification. | 
| *.LastDate | Reports when qualification was last performed (read-only). | 
| *.LastOperator | Reports the name of the user who performed the last qualification (read-only). | 
| *.Interval | Specifies the period of time (in days) between qualifications. | 
| *.WarningPeriod | Specifies the number of days before qualification expiration that Chromeleon will display a reminder that requalification will be necessary. | 
| *.GracePeriod | Indicates for how many more days the user can continue to operate the detector after failing to perform a scheduled qualification. | 
| QualificationDone | Selecting this command after qualification sets *.LastDate to the current date and sets *.LastOperator to the name of the currently logged in user. | 
| UVLampAge | Reports (or resets) the number of hours the deuterium lamp has been on. | 
| VisLampAge | Reports (or resets) the number of hours the tungsten lamp has been on. | 
| VisLampIntensity | Reports the tungsten lamp's intensity at 520 nm. Reported in counts per second. | 
The Administrator Help section provides information about how to install the PDA, PDA-100, or PDA-3000 detector; refer to Hardware Installation:
 ICS-Series Photodiode Array Detector: Overview
 ICS-Series Photodiode Array Detector: Overview
 PDA-100 (DX-LAN) Detector: Overview
 PDA-100 (DX-LAN) Detector: Overview
 Photodiode Array Detector (PDA-100 (USB)/PDA-3000): Installation.
 Photodiode Array Detector (PDA-100 (USB)/PDA-3000): Installation.