Commands and Tips for Third-Party Devices
 Perkin Elmer Autosystem (XL) and Clarus GCs: Detectors
 Perkin Elmer Autosystem (XL) and Clarus GCs: Detectors
The following commands and properties are available for detector control (please note that the display Filter level determines which commands and properties are displayed):
| Property | Range/Values | Description | 
| Attenuation | 1, 2, 4, 8, 16, 32, 64, (and 128, 256, … 65536 if Output is set to Recorder in the Server Configuration program)) | Sets the detector attenuation.² | 
| AZCompensation | On or Off | Enables or disables the Autozero option.² | 
| DetectorStatus | Clarus 400/500 and Clarus 480/580: NotInstalled, Ok, FlameIgnition, FlameOut, BackgroundComp, FilamentOff, Unknown Clarus 600/680: NotInstalled, Ok, FlameIgnition, FlameOut, FlameCheck, BackgroundComp, FilamentOff, IgnitionFail, Unknown | Reports additional information about the status of the detector, especially about the status of the flame. If the flame is out, an error message is displayed. Note: To monitor the temperature of the detector, please refer to property Status below. | 
| DetectorType | Output, FID, ECD, | Indicates the detector type (read-only). Select the detector type in the Server Configuration program: FID: Flame Ionization Detector | 
| FilterTimeConstant | 50 ms (not for the ECD), 200ms, 800ms | Specifies the time constant for the noise filter of the related detector. | 
| FlameThreshold | 0.00-9.99 mV | Flame threshold for evaluating whether the detector flame is lit. If the signal exceeds this value, the GC assumes that the flame is lit. Otherwise, the flame is ignited automatically. | 
| MaxAutoStep | 0.1 to 5.1 s | Specifies the maximum step for the auto step mode. The default setting is 5.1 s. | 
| Range | 1, 20 (FID, NPD) 1 (ECD) -4 - 4 (TCD) | Sets the detector output range.² | 
| Ready | Ready or NotReady | Indicates whether the detector is ready (read-only; level: Advanced or higher). If a PPC is installed, the status of the gas flows is taken into account, also. | 
| Retention | 0 - 10^8 min | Indicates the retention time. | 
| Signal | -10000 - 10000mV | Indicates the current signal value (read-only). | 
| Status | None, Ready, NotReady, FilamentOver, Unknown | Indicates the detector status (read-only). | 
| Step | 0.01 to 4.80 s | Sets the step. The default setting is Auto (recommended). With this setting, the optimum step is selected dynamically. | 
| TempCtrl | On/Off | Turns the heating component of the detector on or off. | 
| Temperature | 100 to 450°C 100 to 350°C (TCD) | Properties related to the detector temperature. 
 If a temperature is set, TempCtrl is automatically set to On. | 
| Temperature. | Sets the nominal temperature. | |
| Temperature. | Indicates the current temperature (read-only). | 
² If the value is entered manually or at time t=0, the value used at the beginning of the run is adapted.
 Tip:
 Tip: 
If a value is entered at another time, this value remains valid until the next setting or the beginning of the next GC run.
In addition, Chromeleon supports the following commands and/or properties if the Autosystem GC includes a PPC:
| Property | Values | Description | 
| <Gasname> | 0–500 ml/min | <Gasname> is the gas specified on the Detectors tab page in the Server Configuration program, e.g., He, N2, H2, Ar/CH4, or Air. 
 You can enter the gas name on the Detectors tab page in the Server Configuration program. | 
| <Gasname>. | Sets the nominal flow for the associated gas. | |
| <Gasname>. | Indicates the current flow for the associated gas (read-only; level: Advanced or higher). | |
| <GasName>_Ctrl | On/Off | 
 | 
| <GasName>_GasType | He, N2, H2, Ar/CH4, Air | Indicates the gas type in the PPC module (read-only; level: Advanced or higher). Note: The gas type is set in the Server Configuration program. | 
| <GasName>_Status | NotUsed, Ready, NotReady, Alarm | Indicates the current status in the PPC module (read-only; level: Advanced or higher). | 
For installation details, refer to: