The following commands and properties are available for inlet control (please note that the display Filter level determines which commands and properties are displayed):
| Property | Range/Values | Description | 
| FollowOven | On or Off | (Only for PPS and POC injectors) If the property is set to On, the inlet temperature follows the oven temperature (+5°C). | 
| InletType | PKD, CAP, GSV, PSSO, PSSI, POCI, POCO | Indicates the inlet type (read-only): | 
| Ready | Ready/NotReady | Set to Ready when the inlet heating component is ready. (If a PPC is installed, the SplitStatus setting must also be Ready.) | 
| Status | None | Indicates the current status of the associated heating component. | 
| TempCtrl | On or Off | Enables or disables temperature control for the inlet. | 
| Temperature | Without PTR: With PTR: | Properties related to the column oven. 
 If a temperature is set, TempCtrl is automatically set to On. | 
| Temperature. | Sets the nominal temperature. For the PSSI and POCI, this command sets a temperature ramp, instead. A maximum of two inlet temperature ramps are allowed. | |
| Temperature. | Indicates the current temperature (read-only) | 
In addition, Chromeleon supports the following commands and/or properties if the Autosystem GC includes a PPC:
| Property | Values | Description | 
| PurgeCtrlMode | Auto, Fixed | Sets the purge flow compensation mode (level: Advanced or higher). (For more information, refer to the manual for the instrument.) | 
| PurgeFlow | 0.0 – 100.0 ml/min | Sets the split flow offset value to compensate the reduction of the split flow due to septum purge flow. (The property can be set only if PurgeCtrlMode = Fixed.) | 
| SplitCtrlMode | Flow, Ratio | Sets the split control mode. Specifies whether the flow or the flow ratio is controlled. | 
| SplitFlow | 0 – 500 ml/min | Sets the target value for the total split flow. (The property can be set only if SplitCtrlMode = Flow.) | 
| SplitRatio | 1-6000 | Sets the target value for the split ratio. (The property can be set only if SplitCtrlMode = Ratio.) | 
| SplitStatus | NotUsed, Ready, NotReady, Alarm, Unknown | Indicates the current split status. | 
| TotalSplitFlow | 0 – 500 ml/min | Indicates the current total flow (split flow + purge flow; read-only). | 