Shodex Refractive Index Detectors (RI-101, RI-102, RI-104): Commands and Tips
The Shodex RI-101 Refractive Index Detector is available from Dionex as part of the Summit HPLC System whereas the RI-102 and RI-104 detectors are not. However, the commands listed below are also supported for the Shodex RI-102 and RI-104 detectors.
In addition to the commands available for various detector types, such as Step, Average, and Autozero (see  Dionex Detectors), the Shodex RI-101 Refractive Index Detector supports the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
 Dionex Detectors), the Shodex RI-101 Refractive Index Detector supports the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
| Property | Minimum | Maximum | Description | 
| Baseline_Shift | 0 mV | 500 mV | Baseline offset (10 mV increments only). (Display level: Advanced) | 
| Data_Collection_Rate | 0.50 Hz | 10 Hz | Enter the rate [Hz] at which Chromeleon collects digital data from the detector. | 
| DemoFileName | 
 | 
 | Name of the standard demo file (Display level: Expert. The property is read-only.) | 
| Drift | 0 nRIU/h | 9990 nRIU/h | Drift value (Display level: Advanced. The property is read-only.) | 
| FirmwareVersion | "1.06" (Note: This is the version number displayed on the startup screen. Other information is not retrievable.) | Standard firmware version. | |
| Integrator_Range | 125 µRIU/V or 500 µRIU/V | Voltage scale for analog integrator output (display level: Advanced) | |
| LampAge | 0 h | 99999 h | Reports the total operating hours of the lamp. Display level: Advanced. The property is read-only.) | 
| LampVoltage | 0.0V | 9.9V | Lamp voltage (The value should not exceed 4.5V. Display level: Advanced. The property is read-only.) | 
| Noise | 0 nRIU | 9990 nRIU | Noise value (Display level: Advanced. The property is read-only.) | 
| NotReadyCauses | Text string (128 characters) | Explanation of why the device is not ready (read-only). | |
| Polarity | Plus | Minus | Output polarity (Display level: Advanced) | 
| Purge | Off | On | Controls the purge valve. | 
| Recorder_Range | 0.25 µRIU | 512.00 µRIU | Voltage scale for full-scale analog recorder output. Only powers of 2 are allowed. (Display level: Advanced) | 
| Rise_Time | 0.10, 0.25, 0.50,1.00, 1.50, 2.00, 3.00, 6.00 s | Sets the response time of the detector upon a signal change in [s] (see also: Rise Time). | |
| Span | 0 µRIU | 999 µRIU | Validation span (The value should be between 487 and 537 µRIU. Display level: Advanced. The property is read-only.) | 
| TemperatureDelta | -99°C | 99°C | Temperature delta (also see: Temperature.Nominal and Temperature.Value. Display level: Advanced. The property is read-only.) | 
| Temperature.Nominal | 30°C | 50°C | Temperature set point. | 
| Temperature.Value | 0°C | 99°C | Measured temperature. | 
Message: "A communication timeout occurred while waiting for device answer in TaskGetAquData".
Possible error cause:
The error message indicates a communication error which occurs if the refractive index exceeds the maximum RI range. This error occurs with earlier firmware versions (1.09 and below).
Solution:
Upgrade the firmware to 1.10 to solve the problem.
The error occurs when running very long chromatograms (longer than 20 hours).
Solution:
With digital data acquisition the length of an analysis (any one injection) is limited to about 20 hours. Longer runs may be interrupted with a "communication timeout" error. In this case use analog data acquisition, for example, by means of a UCI-100 interface.
Further Information
For information about how to install the Shodex RI-detectors, refer to  Shodex HPLC Refractive Index Detectors (RI-101, RI-102, RI-104).
 Shodex HPLC Refractive Index Detectors (RI-101, RI-102, RI-104).