Commands and Tips for Thermo Scientific Devices
 Thermo Scientific Accela Pump: Emergency Program
 Thermo Scientific Accela Pump: Emergency Program
If a fatal error occurs during a batch run, the batch will be aborted automatically. In this case, an Emergency Program can be started to prevent damage to the system.
You may use, for example, the following commands in the emergency program to reduce the flow:
| ; | Program: Emergency.PGM Reduce flow after abort. | 
| 
 | Pressure.LowerLimit | = | 0 [bar] | 
| 
 | Pressure.UpperLimit | = | 500 [bar] | 
| 
 | 
 | 
 | 
 | 
| 0.000 | Flow | = | 0.5000 [ml/min] | 
| 
 | %B.Value | = | 0.0 [%] | 
| 
 | %C.Value | = | 0.0 [%] | 
| 
 | %D.Value | = | 0.0 [%] | 
| 
 | 
 | 
 | 
 | 
| 0.500 | Flow | = | 0.0000 [ml/min] | 
| 
 | %B.Value | = | 0.0 [%] | 
| 
 | %C.Value | = | 0.0 [%] | 
| 
 | %D.Value | = | 0.0 [%] | 
| 
 | 
 | 
 | 
 | 
| 
 | End | 
 | 
 | 
For usage information also refer to  Specifying an Emergency Program.
 Specifying an Emergency Program.