 Thermo Scientific TriPlus RSH / 100 LS Autosampler: Tips for Sequences
 
 Thermo Scientific TriPlus RSH / 100 LS Autosampler: Tips for Sequences
 Note:
 Note:
Headspace, SPME and ITEX mode are only available for the TriPlus RSH.
Restrictions for Headspace, SPME and GC Liquid sequences
When running an analysis using a TriPlus RSH sampler, the following condition must be met:
The GCCycleTime should exceed the total of the GC's program run time and equilibration time.
Sample Overlap (HS, SPME)
Overlapping sample preparation (see Autosampler 
 Control  Overlapping Samples) is supported for headspace and SPME (Solid 
 Phase Microextraction) injections and can be viewed on the Control 
 Panel of the TriPlus RSH Autosampler.
 
 Overlapping Samples) is supported for headspace and SPME (Solid 
 Phase Microextraction) injections and can be viewed on the Control 
 Panel of the TriPlus RSH Autosampler.
For optimal use of the sampler's overlapping sample preparation, Chromeleon sequences should meet the following conditions:
Each run can contain only one program. If you want to use a different program, you must set up a new run.
Use consecutive injections with a different position number for each sample vial. Overlapping samples is not possible for multiple injections taken from the same sample vial.
Use the same injection volume for all 
	 injections. It can be set in the sequence or in the Program Wizard 
	 (see The Control Program  The Program Wizard) of the TriPlus RSH (click "Launch 
	 Editor" in the wizard) if the "Extract injection volume 
	 from TriPlus RSH method" option is enabled in the Server 
	 Configuration program.
 
	 The Program Wizard) of the TriPlus RSH (click "Launch 
	 Editor" in the wizard) if the "Extract injection volume 
	 from TriPlus RSH method" option is enabled in the Server 
	 Configuration program.
 Note:
 Note:
Successful overlapping also depends on a number of other factors, for example:
Analysis 
	 Time settings (see  General 
	 (GC Headspace) and
 General 
	 (GC Headspace) and  Parameters 
	 (SPME) in the Program Wizard).
 Parameters 
	 (SPME) in the Program Wizard).
Overall length of preparations, such as number of washes, incubation time, etc.
The Chromeleon sequence is split into several runs if:
The Chromeleon sequence contains more than one program.
The sample position numbers do not differ from each other.
The injection volumes are different.
Injection Volumes
When specifying injection volumes in the sequence, use the following recommended settings to prevent injection volume rounding:
For HS (GC Headspace) and LS (GC Liquids) modes, set injection volumes to be a multiple of 0.1% of the syringe size. For example, if the syringe size is 2500 µl, set injection volumes to be a multiple of 2.5 µl.
For ITEX (In Tube Extraction) mode, set injection volumes to be a multiple of 1 µl.
 Note:
 Note:
If an injection volume specified in the sequence does not comply with the above recommended settings, it will be rounded to the nearest allowed value and a warning containing the details will be written to the audit trail.
For installation details, refer to Hardware Installation: