Commands and Tips for Thermo Scientific Devices
 Finnigan/TSP: FL2000 and FL3000 Fluorescence Detectors: Different Names in Firmware and Chromeleon
 Finnigan/TSP: FL2000 and FL3000 Fluorescence Detectors: Different Names in Firmware and Chromeleon
For consistency reasons, the names of some Chromeleon properties and commands are different from the names used in the instrument's firmware:
| Instrument | Chromeleon | Description | 
| Lamp Status: On, Off, Run, Off@End | Lamp: On, Off | The instrument provides two additional lamp modes: Run (The lamp is turned on at the beginning of the sample run; it is turned off at the end of the run.) The device driver does not support these modes because you can control the lamp by the corresponding commands in the PGM File.) | 
| Scan | Scan | This command corresponds to 'Scan' entry on the instrument's COMMANDS menu (manual scanning). | 
| Scan Zero | ScanAutozero | The Chromeleon command corresponds to the 'Scan Zero' command on the instrument's COMMANDS menu (manual baseline scanning). The first ScanAutozero command that is included in the PGM File at a non-negative time is not executed immediately. The time is downloaded to the instrument as Autozero Time parameter. The instrument performs the Autozero command at this time. | 
| Autozero Time | Supported via the ScanAutozero command (see above) | 
 | 
| Auto Spectra: On, Off | 
 | Chromeleon sets this parameter always to Off because Chromeleon can perform automatic scanning more reliably by using Trigger Commands. | 
| Zero on l Change | AutozeroOnWlChange | 
 | 
| Detection Type: Fluor Phos | DetectionMode: | In addition, it is possible to measure the Chemiluminescence. Use one of the two modes and turn off the lamp. | 
 Tip:
 Tip:  
After a Reset command, it takes 90 s until the detector is ready to execute another command.
Also, refer to  Finnigan/TSP: FL2000 and FL3000 Fluorescence Detectors
 Finnigan/TSP: FL2000 and FL3000 Fluorescence Detectors