Commands and Tips for Thermo Scientific Devices
 Finnigan/TSP UV3000 Detector
 Finnigan/TSP UV3000 Detector
The following commands and properties are available (please note that the display Filter level determines which commands and properties are displayed):
| Property | Min | Max | Default | Description | 
| AcquisitionMode | Single | Multi | N/A | Changes the data acquisition mode. | 
| Connected | Discon- | Connected | N/A | Indicates whether the detector is connected. | 
| UV_Lamp | Off | On | N/A | Turns the deuterium lamp on or off (l = 190 - 365 nm) | 
| UV_Lamp_ | Intensity_ | Intensity_ | N/A | Indicates whether the deuterium lamp intensity is too low or OK (read-only) | 
| Visible_Lamp | Off | On | N/A | Turns the tungsten lamp on or off (l = 366 - 800 nm) | 
| Visible_Lamp_ | Intensity_ | Intensity_ | N/A | Indicates whether the tungsten lamp intensity is too low or OK (read-only) | 
Channel-specific Commands and Properties
| Property | Min | Max | Default | Description | 
| Response | 0.0 | 9.9 | 1.0 | Time constant. Indicates the time needed by the detector to reach 63% of the full-scale deflection. | 
| 190 nm | 800 nm | 254 nm | Wavelength at which the detector measures the signal. | 
All other channel properties and commands correspond to the standard detector commands (refer to Practical Tips for Device Control  Detector Commands).
 Detector Commands).
 Notes:
 Notes:
The Range property provided by the TSP UV3000 (analog data acquisition) driver is not required for digital data acquisition; thus, it is not available with the TSP UV3000 (digital data acquisition) driver.
Due to extensive communication during the Connect procedure, it may take up to 1 min until the instrument is finally connected. The corresponding message is logged in the Audit Trail.
 Tip:
 Tip:  
The device driver does not support the Scan mode of the detector. Therefore, it is not possible to acquire single spectra.