Commands and Tips for Third-Party Devices
 Varian CP-3800 GC: General
 Varian CP-3800 GC: General
In addition to the  General Device Commands, the Varian CP-3800 GC supports the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
 General Device Commands, the Varian CP-3800 GC supports the following commands and properties (please note that the display Filter level determines which commands and properties are displayed):
| Property | Range/Value | Description | |
| AtmosphericPressure˛ | 0.000 - 2.067 bar | Indicates the atmospheric pressure | |
| Cryo | On/Off | (If cryo cooling is installed) Enables or disables cyro cooling. | |
| CryoHasTimeout˛ | Yes, No | (If cryo cooling is installed) = Yes if the temperature zone did not reach the target temperature, despite Cryo. | |
| CryoTimeout | 0.01 - 999.99 min | (If cryo cooling is installed) Time that the GC waits before turning off the cooling if no sample has been injected. | |
| CryoThreshold | 30-450°C | (If cryo cooling is installed) Temperature above which the GC shall use cryo for cooling. | |
| EndRunTime˛ | 0.01 - 999.99 min | Actual run time of the current program. | |
| Fault˛ | Fault, OK | Indicates whether a fault occurred in the GC*. | |
| OvenFault˛ | Fault, OK | Indicates whether a fault occurred in the column oven. | |
| OvenPower˛ | On, Off | Indicates whether the column oven is on. | |
| OvenReady˛ | Ready, NotReady | Indicates whether the column oven has reached the target temperature and is ready. | |
| ProtocolVersion | n/a | n/a | Protocol version of the instrument (must be '7'). | 
| RunState | NotReady, Equilibrating, Stabilizing, Monitoring, Ready, Running | Indicates the status of the GC. | |
| RunTime | 0.01 - 999.99 min | Indicates the actual retention time during sample processing. | |
| StabilizationTime˛ | 0.00 - 10.00 min | Time that the GC waits after reaching the target temperature. | |
| Temperature | 
 | 
 | Properties related to the column oven temperature. | 
| Temperature. | -99°C - 450°C | Current temperature. | |
| Temperature. | -99°C - UpperLimit | Sets the nominal temperature. | |
| Temperature. | 50 - 450°C | The system aborts the sample batch and starts emergency handling if the temperature exceeds the upper limit. | |
 Note:
 Note:  
˛ These properties are read-only.
Control with or without a non-controlled autosampler:
To synchronize the injection command of the GC with the start of data acquisition, enter the following command instead of the Inject command:
Wait GC.RunState=Running
This applies to injections via a headspace sampler or a valve.
For installation details, refer to Hardware Installation: Installing and Controlling Third-Party Devices  Varian CP-3800 GC: Installation in the Administrator Help section.
 Varian CP-3800 GC: Installation in the Administrator Help section.