EmWavelength
| Instrument Type: | Fluorescence detector | 
| Type: | Integer | 
| Value Range: | Detector-dependent | 
| Default: | Detector-dependent | 
| Related Commands: | |
| Description: | The EmWavelength parameter sets the measuring wavelength. To be able to measure the light emitted by the substance, set the fluorescence detector exactly to the corresponding wavelength. | 
| Function: | If a substance is fluorescent, it emits light at a specific wavelength range. You can measure the Fluorescence with a fluorescence detector that is set to the respective emission wavelength maximum. Select the EmWavelength parameter to enter the emission wavelength. | 
For an example program, refer to  Determining the Optimum Emission Wavelength (RF2000).
 Determining the Optimum Emission Wavelength (RF2000).