Mix
| Instrument Type: | |
| Related Commands: | |
| Description: | The Mix command causes the autosampler to mix defined injection volumes (see Volume) from several sample vials (see Position) in a different sample vial. | 
| Function: | When the action is completed, the autosampler returns the Sampler.Ready signal. Example (ASI-100 Series): | 
| 
 | 0.000 PrepVial BA1 0.000 PrepVolume 10 0.000 Sampler.PrepVial Position + GA1 - RA1 0.000 PrepSubject PrepVial 0.000 Mix 
 | 
| 
 | The above command sequence mixes 10 µl from a vial in the green segment with 10 µl of a vial in the red segment in the corresponding vial of the blue segment. Example (AS/AS50): | 
| 
 | Sampler.Pipet Volume = 20.0, SourceVial = 1, DestinationVial = 10 Sampler.Mix SourceVial = 10, NumberOfTimes = 5, Volume = 30.0 | 
| 
 | In the above command sequence, 20 µl is pipetted from vial 1 and delivered to vial 10. The contents of vial 10 is mixed by drawing in and expelling 30 µl of the vial contents. The mixing cycle is repeated 5 times. You can also use the PrepSpeed (ASI-100) or SyringeSpeed (AS/AS50) command to determine the draw speed of the syringe. | 
| 
 
 | When using highly viscous liquids or liquids with a low boiling temperature, allow the autosampler more time to draw the exact volume exactly without bubbles. | 
For more information, refer to Practical Tips for Device Control: