Installing and Controlling Third-Party 
 Devices
 Gilson 235 and 235p Autosamplers: Installation
 
 Gilson 235 and 235p Autosamplers: Installation
| Device Type: | Autosampler with up to 384 positions | 
| Device Driver: | Gilson: 235 Autoinjector | 
| Supported Hardware Options: | 
 | 
What is required?
| License: | Timebase Class 3 Select About Chromeleon on the Help menu of the Chromeleon client to check whether this license class is available. | 
| Connection: | Modem 
		 cable (Dionex part no. 8914.0128 
		 or 8914.0144A) (For more 
		 information, refer to  | 
| Hardware | A free RS-232 port | 
| Gilson235.pan | 
Hardware Installation
Software Installation
Install the Device Driver in the 
 Chromeleon Server Configuration. (For more information about the program, 
 refer to Software Installation and Communication 
  The 
 Server Configuration 
 Program.)
 The 
 Server Configuration 
 Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the autosampler to the timebase. Select Add Device on the Edit or context menu, and then select Gilson from the Manufacturers list box and 235 Autoinjector from the Devices list box.
The following settings are required:
On the General tab page, select a free COM port and enter the GSIOC address set on the instrument. (The address is displayed on the instrument's rear panel. If necessary, change the address as described above in Device Settings. When selecting the COM ports, it may happen that the Gilson tray editor (required for configuring special sample trays) and the Gilson autosampler configuration software block COM1 and COM2.)
On the Sampler tab page, enter the name of the installed tray file. The pull down box lists all tray files (maximum 9 characters) that are available in the Chromel\Bin directory. If the tray file is located in a different directory, enter the whole path. In addition, you can specify the syringe volume, the maximum syringe speed, and the loop volume. Select the syringe volume from the selection list. This action automatically sets the corresponding maximum syringe speed.
On the Relays tab page, deselect all relays that are not used.
On the Inputs tab page, remove all inputs that are not available.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the Gilson 235 or 235p autosampler, connect the Gilson235.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
Configuring the Sample Tray
The autosampler can access any coordinates. Thus, almost every sample tray including homemade trays can be used. To ensure this flexibility, the driver takes the dimensions and coordinates of the vials from the tray file.
Dionex supplies some tray files for the most commonly used sample racks:
| Tray File | Rack Code | Vial Type | 
| 235_350 | 350 | 96-hole micro titer, flat | 
| 235_350s | 350 | 384-hole micro titer, flat | 
| 235_351 | 351 | 96-hole micro titer, deep | 
| 235_352 | 352 | 2 ml | 
| 235_353 | 353 | 0.7 ml | 
| 235_354 | 354 | 96-hole micro titer, cooled, flat (235p) | 
| 235_354s | 354 | 384-hole micro titer, cooled, flat (235p) | 
If you use a different rack or if special requirements must be fulfilled, you must generate a tray file yourself. For this purpose, use the Tray Editor of the Gilson Software 709. You can also use the editor to modify existing tray files if you wish to number the vials in a different order or reserve special vial positions, for example, for specific applications, by assigning separate zone names.
The program defines the layout of the sample rack and determines the zone names and vial numbering. The vials to be used for injection should be available in the first zone. Name this zone, for example, samples.
 Tip:
 Tip:
With the Inject command, the samples always have to be in the first zone. (The zone number does not influence their actual position.)
You can define additional zones, for example, for fractionating or sample preparation. Please note that the driver only admits the characters A-Z (no umlaut, accents, ß), %, _, and numbers to be used in zone names. Other characters entered in the tray editor will be replaced by _.
Save the result; for example, under \Chromel\bin\MyTray.gty.
Tray Files
Different standard tray files are already available 
 (see below). Install your rack in the autosampler so that the first sample 
 is at the front left. If the corresponding Gilson program is available 
 on your computer, you can freely program different sample arrangements 
 by yourself. For available tray files, refer to  Gilson Autosampler 
 and Fraction Collector: Tray Files.
 Gilson Autosampler 
 and Fraction Collector: Tray Files.
 Tip:
 Tip:
If you change the syringe, communicate the 
 new syringe volume to the autosampler, i.e., make the corresponding setting 
 in the Server Configuration. (For more information about the program, 
 refer to Software Installation and Communication 
  The Server 
 Configuration Program.)
 The Server 
 Configuration Program.)
For specific applications, you can also change the coordinates for waste, rinse station, and injection port.
 Tip:
 Tip:
Dionex recommends using the default values. If these positions are not hit exactly, re-adjust the autosampler.
Injection Modes
For information about the different injection 
 modes, refer to Commands and Tips for 
 Third-Party Devices  Gilson: Injection Modes in the User Help section.
 Gilson: Injection Modes in the User Help section.
Further Information
 Tip:
 Tip:
For more installation details, refer to the manual for the instrument.
For an overview of the different Gilson instruments 
 for which device drivers are available, refer to  Gilson.
 Gilson.