Appendix: Cables and Pin Assignments
 Modem Cable (1:1 Cable)
 
 Modem Cable (1:1 Cable)
Three different Modem Cable types are available for controlling the instruments listed below. The cable to be used depends on the required connectors.
| 
 | Modem Cable | Dionex Part No. | 
| (1) | 9-pin (male) - 9-pin (female) cable (LC Packings part no. 160070; previous LC Packings part no. ULT-S-COM and SW-S-COM) | 8914.0143 | 
| (2) | 25-pin (male) - 9-pin (female) cable (5 m) | 8914.0144A | 
| (3) | 25-pin (male) - 25-pin (male) cable (5 m) | 8914.0128 | 
Instruments for which the cables are required:
Use cable (1) to connect the following instruments:
 Dionex/LC Packings UltiMate 
 System
 Dionex/LC Packings UltiMate 
 System
 Dionex/LC Packings FAMOS Autosampler
 Dionex/LC Packings FAMOS Autosampler
 Dionex/LC Packings Switchos Valve-Switching 
 Unit
 Dionex/LC Packings Switchos Valve-Switching 
 Unit
 ERC RefractoMax 520 Series RI Detectors
 ERC RefractoMax 520 Series RI Detectors
 Polymer Laboratories ELS 2100 (Ice) Detector
 Polymer Laboratories ELS 2100 (Ice) Detector
 Shodex Refractive Index Detectors (RI-101, 
 RI-102, RI-104)
 Shodex Refractive Index Detectors (RI-101, 
 RI-102, RI-104)
 Teledyne Isco Foxy R1 Fraction Collector
 Teledyne Isco Foxy R1 Fraction Collector
 Varian 380-LC/385-LC ELS Detector
 Varian 380-LC/385-LC ELS Detector
Use cables (2) and (3) to connect the following devices:
 CTC Analytics A200S/Fisons AS200
 CTC Analytics A200S/Fisons AS200
 Finnigan/TSP UV3000/Linear UV206 
 Detector
 Finnigan/TSP UV3000/Linear UV206 
 Detector
 Gilson 152 and 156 UV/VIS Detectors
 Gilson 152 and 156 UV/VIS Detectors
 Gilson 235 and 235p Autosamplers
 Gilson 235 and 235p Autosamplers
 Gilson XL Autosampler Series (221XL, 
 222XL, 231XL to 233XL, ASPEC XL)
 Gilson XL Autosampler Series (221XL, 
 222XL, 231XL to 233XL, ASPEC XL)
 Gilson Valves (ValveMate/ValveMate 
 II only)
 Gilson Valves (ValveMate/ValveMate 
 II only)
 Teledyne Isco Foxy 200 Fraction Collector
 Teledyne Isco Foxy 200 Fraction Collector
 Teledyne Isco Foxy Jr Fraction Collector
 Teledyne Isco Foxy Jr Fraction Collector
Pin Assignment
(Also, refer to  Connecting Devices via RS-232):
 Connecting Devices via RS-232):
| Instrument | 
 | PC | ||
| 9-pin | 25-pin 
		  | 
 | 25-pin 
		  | 9-pin 
		  | 
| 1 | 8 | ----------- | 8 | 1 | 
| 2 | 3 | ----------- | 3 | 2 | 
| 3 | 2 | ----------- | 2 | 3 | 
| 4 | 20 | ----------- | 20 | 4 | 
| 5 | 7 | ----------- | 7 | 5 | 
| 6 | 6 | ----------- | 6 | 6 | 
| 7 | 4 | ----------- | 4 | 7 | 
| 8 | 5 | ----------- | 5 | 8 | 
| 9 | 22 | ----------- | 22 | 9 | 
 Notes:
 Notes: 
The pin mapping for connectors with identical pin numbers is 1:1. That is why modem cables are also called 1:1 RS cables.
On the cable (3) with two 25-pin connectors, the remaining 16 pins are also mapped 1:1.
 Tip:
 Tip:
Do not confuse non-crossover modem cables 
 (= 1:1 RS cables) with the crossover  Null Modem 
 Cable!
 Null Modem 
 Cable!
For information about other connections, refer 
 to the installation instructions for the device. (Also, refer to  Installing and Controlling Third-Party 
 Devices.)
 Installing and Controlling Third-Party 
 Devices.)