 CTC Analytics A200S GC Autosampler: Installation
 
 CTC Analytics A200S GC Autosampler: Installation
| Device Type: | Autosampler | 
| Device Driver: | CTC: CTC_A200S_Sampler | 
| Supported Hardware Options: 
 | No special options available. 
 The A200S GC Autosampler is identical with the Fisons AS200 Autosampler. | 
What is required?
Hardware Installation
Software Installation
Install the Device Driver in the 
 Chromeleon Server Configuration. (For more information about the program, 
 refer to Software Installation and Communication 
  The 
 Server Configuration 
 Program.)
 The 
 Server Configuration 
 Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the autosampler to the timebase. Select Add Device on the Edit or context menu, and then select CTC from the Manufacturers list box and CTC_A200S_Sampler from the Devices list box.
The following settings are required:
On the CTC_A200S_Sampler tab page, verify that the Demo Mode is disabled.
On the Sampler tab page, select a free COM port.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the A200S GC autosampler connect the CTCA200S.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
Nothing particular known.
Further Information
 Tip:
 Tip:
For more installation details, refer to the manual for the instrument.
For an overview of the different CTC instruments 
 for which device drivers are available in Chromeleon, refer to  CTC Analytics.