Installing and Controlling Third-Party 
 Devices
 Agilent HP 1050 HPLC System: Installation
 
 Agilent HP 1050 HPLC System: Installation
| Device Type: | HPLC System | ||
| Device Drivers: | Gradient Pump: Agilent or HP: 1050 Gradient Pump (supports four eluents and two relays) Autosampler: Agilent or HP: 1050 Autosampler (supports two relays) UV detector: Agilent or HP: 1050 UV VWD Detector | ||
| Hardware 
		 Options | Device | Agilent Part No. | Validated with firmware version | 
| Gradient pump | HP 79852 | 3.2 | |
| 
 | Autosampler | HP 79855 | 4.2 | 
| 
 | Single-channel | HP 79853A or | 3.02 or | 
What is required?
Hardware Installation
Software Installation
Install the Device Drivers for the 
 Agilent (or HP) 1050 HPLC System in the Chromeleon Server Configuration. 
 (For more information about the program, refer to Software 
 Installation and Communication  The 
 Server Configuration 
 Program.)
 The 
 Server Configuration 
 Program.)
Start the Chromeleon server.
Start the Server Configuration program.
Add the HP 1050 Gradient pump to the timebase. Select Add Device on the Edit or context menu, and then select Agilent (or HP) from the Manufacturers list box and 1050 Gradient Pump from the Devices list box.
The following settings are required:
On the General tab page, select a free COM port.
On the Solvents tab page, specify the number and names of the eluents that are delivered by the pump.
On the Relays tab page, check a box to enable or disable the corresponding relay.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
Follow step 3 above to add the Agilent (or HP) 1050 Autosampler to the same timebase.
The following settings are required:
On the General tab page, select a free COM port.
On the Rack tab page, select the check box of the used rack.
On the Relays tab page, check a box to enable or disable the corresponding relay.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
In addition, add the Agilent (or HP) 1050 UV VWD Detector to the timebase.
The following settings are required:
On the General tab page, select a free COM port.
On the Signals tab page, accept the enabled channel unless you do not want to record a signal.
The Error Levels tab page classifies the severity of any errors that occur. It is generally not necessary to change the default settings.
To control the HP 1050 HPLC system connect the HP1050S.pan Control Panel in the Chromeleon client to the corresponding timebase.
Application
Nothing particular known.
The User Help 
 section provides information about the different commands that the UV 
 detector of the 1050 HPLC System supports; refer to Commands 
 and Tips for Third-Party Devices  Agilent 
 1050 HPLC System: UV Detector.
 Agilent 
 1050 HPLC System: UV Detector.
Troubleshooting
If errors occur during operation, check whether a START/STOP event cable is installed. Remove the cable.
Also, refer to Commands 
 and Tips for Third-Party Devices  GPIB-Connected 
 Devices: Troubleshooting in the User 
 Help section.
 GPIB-Connected 
 Devices: Troubleshooting in the User 
 Help section.
Further Information
 Note:
 Note:
The Thermo Scientific device drivers for 
 HP 1050 HPLC System have been tested successfully with the following firmware 
 versions:  
HP 1050 GP rev 3.2 and HP 1050 S rev 4.2.
If you have problems with firmware versions lower than the ones indicated, a firmware update might help.
 Tip:
 Tip:
For more installation details, refer to the manual for the instrument.
For an overview of the different Agilent/HP 
 instruments for which device drivers are available in Chromeleon, refer 
 to  Agilent (formerly HP).
 Agilent (formerly HP).