Installing and Controlling Thermo 
 Scientific Devices
 TRACE 1300 Series GC: Installation
 
 TRACE 1300 Series GC: Installation
| Device Type | Gas chromatograph | 
| Device Driver: | Thermo Scientific: TRACE 1300 Series 
		 Driver | 
| Supported Hardware Options: | 
 | 
| Other Supported Options: | 
 | 
| 
 
 
 The following options are not supported: 
 | 
What is required?
Hardware Installation
| Device Connection: | In order to connect an autosampler, use the appropriate cable for the respective autosampler. (For information about which cable is required, refer to the autosampler manual.) | 
| Device Settings: | The communication settings must be as follows: static IP address (the IP address can be set via the front panel). 
 Using DHCP for address retrieval of the GC is not supported. | 
| PrepRun Timeout | The GC can be started only when it is in Ready To Inject mode. Therefore, select a PrepRun Timeout that is long enough so that the injection takes place within this time. The PrepRun Timeout can be set via the GC's keypad [Config] [Oven] [PR Timeout]. For information about special module settings, refer to the installation instructions provided with the GC. | 
| Analog Output Interface Settings | Analog Output Interface (AOI) settings need to be configured by means of the GC's keypad. To do so, associate an analog output channel with a detector and configure the gain range of the AOI channels. For details, refer to the installation instructions provided with the GC. For more information about setting 
		 up timed events during the prep-run or run time phases, refer 
		 to  | 
Software Installation
In the Chromeleon Server 
	 Configuration, add the Driver 
	 to a timebase. For details, see  Adding, Configuring or Deleting Devices. 
	 After you added the driver to the timebase, the Configuration dialog 
	 or wizard opens automatically.
 
	 Adding, Configuring or Deleting Devices. 
	 After you added the driver to the timebase, the Configuration dialog 
	 or wizard opens automatically.
Enter the IP address of the GC and click the 'Get' button to automatically read the instrument configuration.
Alternatively, select the desired settings on each configuration tab page. For information about the settings on a page, click the Help button on a page or press F1.
Configuration of a Remote Inject Device
When performing injections using an external sampler that is not controlled by Chromeleon or when doing manual injections, a Remote Inject device has to be installed on the timebase (Note: The TRACE 1300 Series GC Driver must be installed on the timebase prior to the Remote Inject driver):
In the Server Configuration program, add a Remote Inject driver to the timebase. (Select Add Device on the Edit or context menu, and then select Generic from the Manufacturers list box and Remote Inject from the Devices list box.)
A dialog box for 
	 the Remote Inject device appears. On the General 
	 tab page, the default device name is InjectValve. You can accept the 
	 default name or enter a different one.
	From the Inject Port drop-down 
	 list, select the inject port of the TRACE 1300 Series GC Driver (TRACE_1300.Run1 
	 for the front inlet or TRACE_1300.Run2 for the back inlet) to which 
	 the sampler is connected.
Configuration of two AS/AI 1310/3000 GC Samplers plus Remote Inject to allow Manual Injection
The following 
 example illustrates the configuration of two AS/AI 1310/3000 GC Samplers 
 with one TRACE 1300 Series GC using two timebases (timebase sharing) for 
 manual injection (For more information about timebase sharing, refer to 
  Thermo Scientific TRACE 1300 Series GC: Dual Operation). 
 Proceed as described above:
 
 Thermo Scientific TRACE 1300 Series GC: Dual Operation). 
 Proceed as described above: 
If two AS/AI 1310/3000 GC Samplers are configured, you have to select an inject port of the TRACE 1300 Series Driver (TRACE_1300.Run1 for the front inlet or TRACE_1300.Run2 for the back inlet) for each autosampler to establish a connection.
Add a Remote Inject driver to the preferred timebase (front or back) and select one of the remaining inject ports of the TRACE 1300 Series Driver (TRACE_1300.Run3 for the front inlet or TRACE_1300.Run4 for the back inlet).
The configuration could look as follows:

Instrument Method Setup Using a GSV Inlet
A Program for a TRACE 1300 GC with a GSV (Gas Sampling Valve) inlet will not properly trigger the start of the run without some special modification.
Before setting up a program for the TRACE 1300 Series GC with a GSV, a Remote Inject device driver has to be installed in the device tree.
 Note:
 Note:
The TRACE 1300 Series driver must be installed on the timebase prior to the installation of the Remote Inject driver.
For more information about the installation of a Remote Inject driver, see Configuration of a Remote Inject Device above.
Setting up a program for a TRACE 1300 GC with 
 a GSV requires some special modification of the program. For more information, 
 refer to  Thermo Scientific 
 TRACE 1300 Series GC: Program Setup with GSV Inlet.
 Thermo Scientific 
 TRACE 1300 Series GC: Program Setup with GSV Inlet.
Leak Check Control
You can detect gas leaks from an inlet using the Leak Check functionality in Chromeleon. When you perform a leak check, the GC measures the column flow with a true mass flow sensor and compares it to a calculated flow value obtained from the original column constant to see if the numbers match. The instrument detects a gas leak if there is a significant difference between the two values.
 Note:
 Note:
With a TRACE 1310 GC, you can also perform a leak check using the front panel of the GC. Refer to the TRACE 1310 Hardware Manual for details.
To perform a leak check using Chromeleon
 Note:
 Note:
This procedure requires use of the following:
The column-flowmeter connector. Refer to the TRACE 1300/1310 Hardware Manual for details.
Chromeleon's LeakCheck commands, which are available for 
	 the GC FrontColumn/BackColumn 
	 in the Commands dialog box (see  Commands). 
	 To open the Commands dialog box from a control panel, select 
	 Commands on the Control 
	 menu, or press F8.
 
	 Commands). 
	 To open the Commands dialog box from a control panel, select 
	 Commands on the Control 
	 menu, or press F8.
Before starting a leak check, make sure there is a gas flow through the inlet. For example, in the Commands dialog box, set the FlowMode to PressureCtrl, set the PressureCtrl to On, and specify a nominal pressure value: Pressure.Nominal.
Use the following commands to set up the leak check:
LeakCheckPressureSet – Set the pressure for the leak check.
AllowedPressureDrop – Set a value by which the pressure is allowed to drop and still be considered to pass. Note that the available range depends on the setting for LeakCheckPressureSet.
LeakCheckTimeout – Set the duration of the leak check test.
Use the following command to start the leak check:
LeakCheckCtrl
Use the following commands to monitor the leak check:
LeakCheckElapsedTime – Reports the elapsed time for the leak check.
LeakCheckRemainingTime – Reports the remaining time for the leak check.
Finally, use the following command to see the result of the leak check:
LeakCheckStatus – Reports the status of the leak check.
When the leak check is started, the system is automatically pressurized with carrier gas and sealed to perform a leak check. At the end of the leak check, if the system is free of leaks, the Leak Check Passed message is displayed. If any leaks are found, an error message is displayed; in this case, eliminate the leaks and repeat the leak check procedure.
 Note:
 Note:
Only a previous Column Characterization, performed under leak-free conditions, can ensure the validity of the subsequent leak check responses.
You cannot start a leak check while a leak check or a column evaluation is already running on the same or another column.
Further Information
For detailed installation instructions, refer to the operating instructions for the instrument.
For an overview of the different Thermo Scientific/Thermo 
 Finnigan/ThermoQuest/TSP instruments for which device drivers are available 
 in Chromeleon, refer to  Thermo Scientific.
 Thermo Scientific.